Abstract:
A mechanism model of enzyme injection biosensor was established and the accuracy of the model was verified through experiments in this study. The parameters of the reaction model were determined by fitting the voltage data with model and the voltage data was obtained by detecting 1 and 2 mg mL(-1) glucose solution. Accuracy of the model was verified by comparing the fitting curve and the actual curve. The fitting curve was obtained by taking the value of 3 mg mL(-1) into the model and the actual data was detected by actual sensors. Experimental results showed that the liquid enzyme Michaelis constant K-m was 1.97, the correlation coefficient between mathematical model and the actual sensor working model was 0.998.