Measurement of material elastic constants for limited-size samples by ultrasonic method is described and validated. Based on acoustic microscope technology, the material elastic constants are determined by longitudinal wave and leak surface wave velocities simultaneously measured by developed ultrasonic system with a line-focus PVDF transducer. In this paper, the elastic constants of limited-size Cr(2)O(3) doping on 0.2 PZN-0.8 PZT piezoelectric ceramic wafer sample are determined and the measurement errors are analyzed. The experimental results show this ultrasonic system can be used for measurement of material elastic constants for limited-size samples with high measurement precision, and the relative errors for Poisson's ratio and Young's module measurements are, respectively, less than 1% and 3%. It can satisfy the requirement of engineering and science research. Furthermore, it is suitable to measure elastic constants of both isotropic and anisotropic materials by ultrasonic method with help of developed ultrasonic system. (C) 2009 Elsevier Ltd. All rights reserved.