Background: The STM CP-24 centrifugal pump is a newly developed centrifugal pump for extracorporeal membrane oxygenation equipment. This study aimed to combine hydraulic experiments, hemodynamic numerical simulations, and standard in vitro hemolysis experiments to investigate the comprehensive performance of this centrifugal pump. Methods: In vitro experiments were first done to obtain the pressure-flow data of the centrifugal pump in its working range to evaluate its hydraulic performance. Next, the commonly used clinical working points were selected as boundary conditions, and a computational fluid dynamics method was applied to evaluate its hemodynamic performance. The blood pressure distribution, blood flow fields, and high-wall-shear-stress zones in the centrifugal pump were determined as indicators for hemodynamic evaluation. Finally, standard in vitro hemolysis experiments were performed to test the blood compatibility of this centrifugal pump (n=3 blood samples). In addition, its blood compatibility was evaluated in the form of the normalized index of hemolysis (NIH). Results: The pressure-flow curve of the centrifugal pump showed that the head pressure and flow of the centrifugal pump showed a mostly linear relationship within the whole working range. When the rotation speed of the centrifugal pump was 5,500 rpm, it achieved a hydraulic performance of 550 mmHg head pressure and 8 L/min output flow, which could meet the clinical needs of extracorporeal membrane oxygenation. Analysis of computational fluid dynamics data indicated that the centrifugal pump had excellent hemodynamic performance: even distribution of blood pressure in the pump, no blood flow stagnation zone or dead zone in the overall flow field, and secondary flows in the gap between the rotor and the volute that significantly reduced the volume of the low-blood-flow zone close to the impeller. There was no obvious high-shear-stress zone on the surface of the volute or the impeller, which will effectively reduce the risk of thrombosis. In vitro hemolysis experiments indicated that the centrifugal pump had excellent blood biocompatibility, with a NIH =0.0125?0.0022 g/100 L. Conclusions: The STM CP-24 centrifugal pump has excellent hydraulic performance, a reasonable design of the hemodynamic structure of the blood pump, and excellent blood compatibility. Therefore, it can meet Background: The STM CP-24 centrifugal pump is a newly developed centrifugal pump for extracorporeal membrane oxygenation equipment. This study aimed to combine hydraulic experiments, hemodynamic numerical simulations, and standard in vitro hemolysis experiments to investigate the comprehensive performance of this centrifugal pump. Methods: In vitro experiments were first done to obtain the pressure-flow data of the centrifugal pump in its working range to evaluate its hydraulic performance. Next, the commonly used clinical working points were selected as boundary conditions, and a computational fluid dynamics method was applied to evaluate its hemodynamic performance. The blood pressure distribution, blood flow fields, and high-wall-shear-stress zones in the centrifugal pump were determined as indicators for hemodynamic evaluation. Finally, standard in vitro hemolysis experiments were performed to test the blood compatibility of this centrifugal pump (n=3 blood samples). In addition, its blood compatibility was evaluated in the form of the normalized index of hemolysis (NIH). Results: The pressure-flow curve of the centrifugal pump showed that the head pressure and flow of the centrifugal pump showed a mostly linear relationship within the whole working range. When the rotation speed of the centrifugal pump was 5,500 rpm, it achieved a hydraulic performance of 550 mmHg head pressure and 8 L/min output flow, which could meet the clinical needs of extracorporeal membrane oxygenation. Analysis of computational fluid dynamics data indicated that the centrifugal pump had excellent hemodynamic performance: even distribution of blood pressure in the pump, no blood flow stagnation zone or dead zone in the overall flow field, and secondary flows in the gap between the rotor and the volute that significantly reduced the volume of the low-blood-flow zone close to the impeller. There was no obvious high-shear-stress zone on the surface of the volute or the impeller, which will effectively reduce the risk of thrombosis. In vitro hemolysis experiments indicated that the centrifugal pump had excellent blood biocompatibility, with a NIH =0.0125 +/- 0.0022 g/100 L. Conclusions: The STM CP-24 centrifugal pump has excellent hydraulic performance, a reasonable design of the hemodynamic structure of the blood pump, and excellent blood compatibility. Therefore, it can meet clinical needs.

关键词:

Centrifugal pump hemolysis hemodynamics normalized index of hemolysis (NIH)

作者机构:

  • [ 1 ] [Fu, Minrui]Beijing Univ Technol, Fac Environm & Life, Beijing 100124, Peoples R China
  • [ 2 ] [Wang, Weining]Beijing Univ Technol, Fac Environm & Life, Beijing 100124, Peoples R China
  • [ 3 ] [Gao, Bin]Beijing Univ Technol, Fac Environm & Life, Beijing 100124, Peoples R China
  • [ 4 ] [Liu, Youjun]Beijing Univ Technol, Fac Environm & Life, Beijing 100124, Peoples R China
  • [ 5 ] [Liu, Gang]Chinese Acad Med Sci & Peking Union Med Coll, Fuwai Hosp, Natl Ctr Cardiovasc Dis, State Key Lab Cardiovasc Dis, Beijing 100037, Peoples R China
  • [ 6 ] [Ji, Bingyang]Chinese Acad Med Sci & Peking Union Med Coll, Fuwai Hosp, Natl Ctr Cardiovasc Dis, State Key Lab Cardiovasc Dis, Beijing 100037, Peoples R China
  • [ 7 ] [Wang, Weining]Jiangsu STMed Technol Co Ltd, Suzhou, Peoples R China
  • [ 8 ] [Chang, Yu]Zhejiang Univ, Childrens Hosp, Sch Med, Natl Clin Res Ctr Child Hlth, Hangzhou, Peoples R China

通讯作者信息:

  • 刘有军

    [Gao, Bin]Beijing Univ Technol, Fac Environm & Life, Beijing 100124, Peoples R China;;[Liu, Youjun]Beijing Univ Technol, Fac Environm & Life, Beijing 100124, Peoples R China;;[Ji, Bingyang]Chinese Acad Med Sci & Peking Union Med Coll, Fuwai Hosp, Natl Ctr Cardiovasc Dis, State Key Lab Cardiovasc Dis, Beijing 100037, Peoples R China

查看成果更多字段

相关关键词:

来源 :

ANNALS OF TRANSLATIONAL MEDICINE

ISSN: 2305-5839

年份: 2021

期: 8

卷: 9

JCR分区:3

被引次数:

WoS核心集被引频次: 9

SCOPUS被引频次:

ESI高被引论文在榜: 0 展开所有

万方被引频次:

中文被引频次:

近30日浏览量: 1

归属院系:

在线人数/总访问数:350/4976506
地址:北京工业大学图书馆(北京市朝阳区平乐园100号 邮编:100124) 联系我们:010-67392185
版权所有:北京工业大学图书馆 站点建设与维护:北京爱琴海乐之技术有限公司