Auto-calibration of the amplitude-frequency characteristic of oscilloscope calibrator is realized in the paper. In the system, power meter is controlled through GPIB bus by Windows VC++. The auto-calibration system can automatically provide effective value (or power) and relative error of oscilloscope calibrator output level fast, and create calibration template, preserve calibration data, and generate calibration certificate etc. And the auto-calibration system can also calibrate common sine-wave signal source. This auto-calibration system of oscilloscope calibrator is already used in automatic calibration test system of "Pulse Parameters standard device" in National Institute of Metrology, which has improved the efficiency of calibrating oscilloscope calibrator greatly.
关键词:
通讯作者信息:
电子邮件地址:
来源 :
PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1-2
年份: 2008
页码: 1134-1137