• 综合
  • 标题
  • 关键词
  • 摘要
  • 学者
  • 期刊-刊名
  • 期刊-ISSN
  • 会议名称
搜索

作者:

Zhu, Jiang-Miao (Zhu, Jiang-Miao.) | Zhao, Lin-Xiao (Zhao, Lin-Xiao.) | Miao, Jing-Yuan (Miao, Jing-Yuan.) | Yang, Jie (Yang, Jie.)

收录:

EI Scopus PKU CSCD

摘要:

This paper constructed a calibration system of oscilloscope based on the optical pulse source. The 70GHz sampling module 86118A of Agilent 86100C was calibrated based on optical pulse source using the standard pulse method. In the calibration process, the influences on calibration results of three kinds of time base errors were researched for the first time, e. g., time base drift was removed using a cross-correlation algorithm, time base distortion was corrected using the least squares method, and jitter was removed by using PDF deconvolution. Experiment results showed that calibration of 70GHz broadband oscilloscope was realized, the time base errors could be effectively removed by this method, and the accuracy of the calibration was satisfied. © 2018, Chinese Institute of Electronics. All right reserved.

关键词:

Calibration Cathode ray oscilloscopes Deconvolution Least squares approximations Optical correlation Stress intensity factors

作者机构:

  • [ 1 ] [Zhu, Jiang-Miao]Faculty of Information Technology of Beijing University of Technology, Beijing; 100124, China
  • [ 2 ] [Zhao, Lin-Xiao]Faculty of Information Technology of Beijing University of Technology, Beijing; 100124, China
  • [ 3 ] [Miao, Jing-Yuan]National Institute of Metrology, Beijing; 100013, China
  • [ 4 ] [Yang, Jie]Faculty of Information Technology of Beijing University of Technology, Beijing; 100124, China

通讯作者信息:

电子邮件地址:

查看成果更多字段

相关关键词:

相关文章:

来源 :

Acta Electronica Sinica

ISSN: 0372-2112

年份: 2018

期: 4

卷: 46

页码: 945-951

被引次数:

WoS核心集被引频次: 0

SCOPUS被引频次: 2

ESI高被引论文在榜: 0 展开所有

万方被引频次:

中文被引频次:

近30日浏览量: 3

归属院系:

在线人数/总访问数:142/3610881
地址:北京工业大学图书馆(北京市朝阳区平乐园100号 邮编:100124) 联系我们:010-67392185
版权所有:北京工业大学图书馆 站点建设与维护:北京爱琴海乐之技术有限公司