Author:

Zhao, Ke-Jia (Zhao, Ke-Jia.) | Liu, Ming-Liang (Liu, Ming-Liang.) | Gao, Xiao-Xun (Gao, Xiao-Xun.) | Lu, Feng (Lu, Feng.) | Zhu, Jiang-Miao (Zhu, Jiang-Miao.) | Miao, Jing-Yuan (Miao, Jing-Yuan.)

Indexed by:

EI Scopus PKU CSCD

Abstract:

The nose-to-nose calibration procedure can obtain the impulse response and complex transfer function more accurately, comparing with other technologies to calibrate broadband sampling oscilloscope. Because it should meet the needs to connect the two oscilloscope input channels directly in nose-to-nose calibration procedure, it is very difficult to operate the oscilloscopes through front panel. So an auto-calibrating system of broadband sampling oscilloscope based on nose-to-nose calibration procedure is introduced. The system build-up, system function and calibration results are described in detail. The calibration of this system is implemented on PC.